Model 530 Operating Instructions
Power and Measure Transmitter Mode
Transmitter Input
(Sensor)
(Process Signal)
(Simulated Input)
Receiver
Power Supply
0 to 60 VDC
Typical
2-Wire
Transmitter
+ IN - REF +OUT-
The Model 530 supplies 24 Volts to the transmitter and displays the output in mA or % on the Model 530 display. If the Model 530’s
HART
®
protocol mode is selected, an internal 250 load resistor is connected in series with the output as specified for proper
communication by HART devices. This eliminates the requirement of connecting a separate 250 load resistor in the field as
indicated in field hookup diagrams of HART
®
enabled devices.
Read Mode
Transmitter Input
(Sensor)
(Process Signal)
(Simulated Input)
Receiver
Power Supply
0 to 60 VDC
Typical
2-Wire
Transmitter
+ IN - REF +OUT-
The Model 530 can read loop currents from 0-24 mA. The Model 530 limits current in read mode to less than 24 mA to protect the
devices in the loop from over voltage or over current conditions.